Description
- Pre-warm
mode control, manual mode control and baby mode controlled by micro-computer;
- Set
temperature and baby temperature can be displayed separately;
- Horizon
angle of heater head and inclination of the bassinet can be adjusted;
- The
panels around bassinet can be turned outward;
- X-ray
cassette under the bassinet;
- APGAR
timer;
- Failure
alarm indication;
- RS232
connector;
- Vertical
height adjustment stand and Baby scale is optional.
- Provided with Breath resuscitation bag and Infant head
fixing unit.
- Featured with LED Phototherapy unit;
- Timer
to count phototherapy time.
- �
�Standard configuration
-
The warmer module, bassinet, main column, stand, controller, I.V. Pole and
tray.
- All
of above three models can be configured with vertical height adjustment stand.
�Specification
- Power
supply: AC220-230V/50Hz
- Maximum
Power Output: 700W (fixed stand)
- Mode
of control: Pre-warm mode control, manual mode control and baby mode controlled
by micro-computer
- Baby
mode temperature control range :34.5-37.5?
- Temperature
sensor display range: 5?-65?
- Accuracy
of skin temperature sensor: =0.3?
- Deviation
between the measure by the skin temperature sensor and control temperature: =0.5?
- Temperature
uniformity of mattress: =2?
- Angle
of warmer module: 0�,30�,60�for
two ways;
- Inclination
of the bassinet: �10�
- APGAR
timer: it will give sound when the unit tones at1�, 5 �, 10�;
- Failure
alarm: over temperature alarm, deviation alarm, sensor failure alarm, power
failure alarm, settingalarm, checking alarm and so on.
- Max.
bilirubin radiation on the mattress effective range :=1000uW/cm2(HBHKN-93B,HBHKN-93A)
- Bilirubin
uniformity of radiating on the mattress effective range:>0.4(HBHKN-93A/HBHKN93B)
- Wave
length of blue light: 420-470nm
- �
�Environmental
temperature
- Operating
range: 18?-30?
- Ambient
air movement rate: =0.3m/s
�Package
- Package:
each unit is packed to one carton
- Size
(Main body): 127cm�81cm�112cm
- Gross
Weight: 110kg, 120kg, 142kg
|